three calibration methods
INSTRUCTION VIDEOS FROM OUR IPP YOUTUBE CHANNEL
1 In-Situ 2 Comparison 3 Replication
|
|
|
Using the Insacal to calibrate conductivity Sensors without taking sensors Off Line
|
Using INSACAL to Calibrate Sanitary Conductivity sensors to meet the USP645 requirements
|
Conductivity – the importance of temperature and its compensation - CIP, Conductivity - performance and regulatory requirements
|
For all these methods:
- Calibration against superior master meter EXAxtcal
- Multipoint cell constant calibration possible
- Multipoint temperature calibration possible
- Accurate - 4:1 test accuracy ratio (TAR)
- Independent of test solution
- Robust build and uncertainty as low as 0.24%
For Pharmaceutical and Medical purposes the quality of water, as an excipient has been defined by the European Pharmacopia (EP) as well as the United States Pharmacopia (USP). In order to ship/sell your product in Europe or the United States of America, you must be able to prove that these requirements are met.
For Purified Water and Water for Injection (WFI) the USP defines the following requirements:
• Meter reports uncompensated Conductivity or uncompensated Resistivity.
• The display resolution is 0.1 μS/cm or better.
• The meter reads accurately to ±0.1 μS/cm when a 0.1% precision resistor replaces the sensor (to calibrate/verify the meter).
• The sensor cell constant is calibrated/verifiedto ±2%
• Temperature accurate to 2°C (effective USP 28)
Yokogawa have years of experience in the development of highly accurate conductivity instrumentation for your on-line measurements.
Irish Power and Process and Insatech have jointly developed the Insacal Conductivity Calibrator in order to help you achieve these difficult requirements.
This is the first master Conductivity standard, accredited to 0.4% or better, in order to allow you meet your WFI quality requirements of 2% cell constant verification with a 4:1 TAR. Download The Insacal Conductivity Master Meter for Regulatory Compliance flyer for further information.
For Purified Water and Water for Injection (WFI) the USP defines the following requirements:
• Meter reports uncompensated Conductivity or uncompensated Resistivity.
• The display resolution is 0.1 μS/cm or better.
• The meter reads accurately to ±0.1 μS/cm when a 0.1% precision resistor replaces the sensor (to calibrate/verify the meter).
• The sensor cell constant is calibrated/verifiedto ±2%
• Temperature accurate to 2°C (effective USP 28)
Yokogawa have years of experience in the development of highly accurate conductivity instrumentation for your on-line measurements.
Irish Power and Process and Insatech have jointly developed the Insacal Conductivity Calibrator in order to help you achieve these difficult requirements.
This is the first master Conductivity standard, accredited to 0.4% or better, in order to allow you meet your WFI quality requirements of 2% cell constant verification with a 4:1 TAR. Download The Insacal Conductivity Master Meter for Regulatory Compliance flyer for further information.